Print Back
Capacity: 220 g to 520 g            Readability: 0.1 mg

RADWAG AS X2 Plus Analytical Balances with innovative ergonomic mechanical design and system solutions are the best choice for use in laboratories and pharmaceuticals and advanced-class products.

FREE Freight Offer is for Continental US only via UPS Ground

RADWAG AS X2 PLUS Analytical Balances

Unit of Measure

Specifications

FREIGHT - UPS-GND

FREE

Model

AS 310.X2 PLUS

Type

Analytical

MFR

RADWAG

Capacity

310 g

Readability (d)

0.1 mg

Minimum Load

10 mg

Tare Range

-310 g
5 Percent Maximum Repeatability1 0.07 mg
Maximum Repeatability2 0.1 mg

Minimum Weight United States Pharmacopoeia's (USP)

140 mg

Minimum Weight United States Pharmacopoeia's (USP) (U = 1 %, k = 2)

14 mg

Linearity (±)

0.2 mg

Eccentric Load Deviation

0.3 mg

Stabilization Time

2.5 sec

Internal Calibration

true

Pan Size

Ø 100 mm

Display

5 in. capacitive color touch screen

Keypad

6 Keys

Protection class

IP 43

Databases

7

Touch-free operation

2 programmable proximity sensors

USB-A

2

USB-B

1

Ethernet

10 / 100 Mbit

RS 232

1

DB9

Tare and Print External

Wi-Fi®

802.11 b/g/n

Direct Current (DC) Power Supply

12 to 60 V

Power Consumption

4 W

Operating Temperature

10 to 40 ºC
Atmospheric Humidity3 40 to 80 %

Transport and Storage Temperature

-20 to 50 ºC

Weighing Pan Dimensions

Ø 100 mm

Weighing Chamber Dimensions

190 x 190 x 226 mm

Weighing Device Dimensions

351 x 205.5 x 348 mm

Net Weight

7.3 kg

Gross Weight

9.3 kg20 lb

Specifications

  • Ergonomic Mechanical Design
  • Spacious weighing chamber and large open-door clearance allow easy access to the weighing pan and facilitate use of
  • laboratory glassware of various sizes and dimensions. Improved aluminum base of the balance guarantees stability of the weighing system.
  • DUAL-CLICK system facilitates tool-free disassembly and assembly of the weighing chamber. Locating the USB interface at the balance front makes it easier to communicate with peripherals.
  • The in-built sensors measure tilt and display graphic message on the weighing device screen.
  • Auto Levelling system facilitates adjustment, controls the level state, and informs about level deviations, all this in accordance with GLP and GMP.
  • AS X2 PLUS balances feature numerous functions supporting and supervising mass measurement in laboratory, facilitating operation and limiting the risk of "human factor" errors, e.g., automatic adjustment, data control, ALIBI memory, automatic reports and ambient conditions monitoring.
  • Weighing chamber panes feature antistatic coating compensating electrostatic charges on the sample and accessories used for mass measurement.
  • Antistatic coating has been applied in balances with the readability of d=0.01mg.
  • Functions displayed on the color 5" touch screen can be customized to suit carried out process and user’s needs.
  • Flexible configuration of buttons, labels and text fields is also a great advantage.
  • AS X2 PLUS balances are equipped with Kensington Lock, which allows to secure the device against theft.
Databases in X2 series balances: In new AS X2 series balances the information system is based on 8 databases which allow several users to work with databases storing several products. The registered weighing results can be subject to further analysis. The data is registered in 8 databases:
  • Users (up to 100 users)
  • Products (up to 5,000 products)
  • Weighments (up to 50,000 weighments)
  • Packaging (up to 100 packaging types)
  • Formulas (up to 100 formulas)
  • Clients (up to 100 clients)
  • ALIBI memory (up to 500,000 weighments)
  • Customers (up to 1,000)
Additionally, it is possible to create formulation report (max 500 reports)

Shipping Dimensions and Weight

Shipping Weight / Dimensions

20 lb / 21 x 20 x 17 in.

Features

N/A

  • Auto Leveling
  • 2x USB-A
  • USB-B
  • Ethernet
  • IR Touchless Sensors
  • RS232
  • WI-FI
  • Warranty :2 Year Electronics / 3 Year Mechanism

Applications

N/A

  • ALIBI Memory
  • Ambient Conditions Monitoring
  • Animal Weighing
  • Autotest
  • Checkweighing
  • Density Determination
  • Dosing
  • Formulation
  • GLP Procedures
  • IR Proximity Sensors
  • Mass for Titrator
  • Multilingual menu
  • Newton Unit Measurement
  • Parts Counting
  • Peak Hold
  • Percent Weighing
  • Replaceable Unit
  • Statistics
  • Statistical Quality Control
  • Under-Pan Weighing

Notes

N/A Ship point FL 33179

  • 1 Repeatability is expressed as a standard deviation from 10 weighing cycles.
  • 2 Repeatability is expressed as a standard deviation from 10 weighing cycles.
  • 3 Stabilization time depends on external conditions and dynamics of placing loads on a pan, determined for FAST profile.